The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C,…
icscert·CWE-284·Published 2019-03-25
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product’s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device.
El protocolo de telemetría Conexus utilizado en las versiones 24950 y 24952 de Medtronic MyCareLink Monitor, CareLink Monitor versión 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D e ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD y CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD y Viva CRT- D no implementa autenticación o autorización. Un atacante con acceso adyacente de corto alcance a un producto afectado, en situaciones donde la radio del producto está encendida, puede inyectar, reproducir, modificar o interceptar datos dentro de la comunicación de telemetría. Este protocolo de comunicación proporciona la capacidad de leer y escribir valores de memoria en dispositivos cardíacos implantados afectados; por lo tanto, un atacante podría explotar este protocolo de comunicación para cambiar la memoria en el dispositivo cardíaco implantado.
| Version | Type | Source | Base | Exp | Impact | Vector |
|---|---|---|---|---|---|---|
| 2.0 | Primary | NVD | 3.3 | 6.5 | 2.9 | AV:A/AC:L/Au:N/C:N/I:P/A:N |
| 3.1 | Primary | cve.org | 9.3 | — | — | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H |
| 3.1 | Primary | NVD | 6.5 | 2.8 | 3.6 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N |
| 3.1 | Secondary | NVD | 9.3 | 2.8 | 5.8 | CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:H |