A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.…
icscert·CWE-120·Published 2018-02-15
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
A Classic Buffer Overflow issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump does not verify input buffer size prior to copying, leading to a buffer overflow, allowing remote code execution on the target device. The pump receives the potentially malicious input infrequently and under certain conditions, increasing the difficulty of exploitation.
Se ha descubierto un problema de desbordamiento de búfer clásico en Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, en versiones 1.1, 1.5 y 1.6. Un componente de terceros empleado en la bomba no verifica el tamaño de entrada del búfer antes de la copia, lo que desemboca en un desbordamiento de búfer que permite la ejecución remota de código en el dispositivo objetivo. La bomba recibe las entradas potencialmente maliciosas de forma infrecuente y bajo ciertas condiciones, lo que dificulta su explotación.
| Version | Type | Source | Base | Exp | Impact | Vector |
|---|---|---|---|---|---|---|
| 2.0 | Primary | NVD | 6.8 | 8.6 | 6.4 | AV:N/AC:M/Au:N/C:P/I:P/A:P |
| 3.0 | Primary | NVD | 8.1 | 2.2 | 5.9 | CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H |