Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System,…
icscert·CWE-285·Published 2019-03-25
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Hospira Plum A+ Infusion System versión 13.4 y anteriores, Plum A+3 Infusion System en su versión 13.6 y anteriores y Symbiq Infusion System, en su versión 3.13 y anteriores proporcionan privilegios root a usuarios no autenticados en el Puerto 23/TELNET por defecto. Un usuario no autorizado podría enviar comandos a la bomba. Hospira recomienda con los usuarios cierren el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado también el Plum 360 Infusion System que no es vulnerable a este problema.
| Version | Type | Source | Base | Exp | Impact | Vector |
|---|---|---|---|---|---|---|
| 2.0 | Primary | NVD | 10.0 | 10.0 | 10.0 | AV:N/AC:L/Au:N/C:C/I:C/A:C |
| 3.0 | Primary | NVD | 9.8 | 3.9 | 5.9 | CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |