Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and…
icscert·CWE-312·Published 2019-03-25
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Las claves inalámbricas se almacenan como texto en claro en Hospira Plum A+ Infusion System en su versión 13.4 y anteriores, Plum A+3 Infusion System en su versión 13.6 y anteriores y Symbiq Infusion System, en su versión 3.13 y anteriores. Hospira recomienda con los usuarios cierren el Puerto 20/FTP y el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado también el Plum 360 Infusion System que no es vulnerable a este problema.
| Version | Type | Source | Base | Exp | Impact | Vector |
|---|---|---|---|---|---|---|
| 2.0 | Primary | NVD | 5.0 | 10.0 | 2.9 | AV:N/AC:L/Au:N/C:P/I:N/A:N |
| 3.0 | Primary | NVD | 7.5 | 3.9 | 3.6 | CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N |